DR #27 - Tektronix 2213A Oscilloscope CRT Replacement & Tune-Up

Duality Repair
132 followers
6 Video Views·Mar 12, 2025
In this video, I document the work done on another Tektronix 2213A oscilloscope. This includes my first CRT replacement, a re-cap, and some troubleshooting.
hey everyone welcome to duality repair I
have another Tektronix 22:13 a 60
megahertz oscilloscope to repair this
one's got a pretty interesting problem
and rather than show you I'll let you
listen to it so you can hear that
rattling around that's the CRT glass it
imploded so this is going to be an
interesting one for me this is my first
CRT replacement and this was sent by a
viewer who purchased this knowing that
it had a problem he also purchased a
spare CRT and I verified the models are
the exact same so that will work which
is good obviously I'm gonna replace the
CRT I'm also going to tune this up as
best I can you know I looked around the
boards a little bit and everything looks
to be in really good condition so I
think once we get the CRT replaced we
replace the electrolytic and maybe clean
the potentiometers all the dials are
here and everything I think this thing
is gonna run really well
I'll go through any of the adjustment
procedures that I can that are
highlighted in the service manual and
we'll get this thing ready for ready for
action
so CRT replacement first things first
I'll remove the anode so the anodes up
here just on plugs there and I've
already done this so ordinarily be a
little more safe but I've already done
this you just pull it out here's the
anode tip right there when this is live
it's at it looks like seven eight
thousand volts and even when it's been
off for a while it can still keep a
charge so what you want to do is you
want to discharge this to chassis so you
want to make sure that your hands aren't
on the chassis and the chassis is not
sitting on anything conductive before
you discharge it to the chassis and then
now you know that the anode the CRT is
safe this thing has been off for several
weeks maybe months and it's also broken
and I've also already grounded this to
the chassis so I know it's safe so
that's step one step two is removing the
front of the CRT or disconnecting it and
that's just these two bottom screws here
Torx head t 15 maybe 250 220 and then we
have to remove the vertical and
horizontal deflection
wires so here are the vertical wires
right here you just remove them and then
the horizontal are down here I'm
definitely going to want to label these
so that I know which way where they go
when I put them back maybe T for top and
B for bottom or something like that same
with the bottom the horizontal
deflection wire so make sure to do that
the only thing I'm unsure of and I
haven't consulted the service manual
with that is is how to get the back out
I think once I remove the front two
screws and I get these disconnected and
I get the anode disconnected it might
just slide out to the right if I have a
little trouble consult the service
manual since it does have the CRT
removal procedure in there but I'm gonna
get working on that all right I got the
CRT out it wasn't too much of a hassle
there were a few things that I missed
when I explained earlier number one
these set of pins here you can see the
two pins coming out of this board this
is for the trace rotation so you do have
to disconnect that trace rotation
connector before you can remove the CRT
and then two other things I mentioned
that the front faceplate is connected
with the two screws on the bottom and
the front I missed the two screws on the
top in the back so there's screw here
and there's another screw in this corner
here those also have to be removed once
you get all of that removed and the
other things I mentioned it does just
slide out to the right which is perfect
and so after I cleaned up all of the
glass I'm ready to go now since it's
well never the case we get to look at a
CRT or that I get to anyway I thought
we'd take a look at this huh pretty
awesome so obviously it so got a few
glass fragments on there those are just
still adhered to some of the pins I'm
not going to worry about trying to
remove them obviously normally this
entire enclosure is encased in encased
in glass all the way to the front screen
here also obviously broken it sits like
this it's encased in glass and under
vacuum but let's take a look around
since we really never get to analyze
something like this so the back is the
connector you can see all the pins there
one good thing that I noticed I was
worried about how it was going to
reconnect the new one you do not want to
connect this incorrectly so if you can
see there's a
key right there this little slot it's
keyed so that you can only insert this
into the mating connection one-way so
you can't really install it incorrectly
which is great
foolproof and so a little bit on the CRT
functionality if we start on the left
this little cylinder there that's going
to be the cathode where the electrodes
are generated all of this here is going
to be various focusing so beam focusing
and beam acceleration to focus and
accelerate the electron beam down the
electron gun we get here you can see
this gold metal right here and there's
another one on the opposite end that's
one of the deflection plates so I think
that's the vertical deflection plate
although I'm not sure and then we have
another set of plates right here you can
see these a lot easier these are the
opposite set of plates so either the
vertical or horizontal and so these are
the voltage difference between the two
plates is going to determine the amount
of deflection of the electron beam
before it hits the screen and that's it
for the electron gun portion let's take
a look at the screen portion there's not
too much to look at so you can see we've
got this coil here and that's connected
to this trace rotation connector that I
mentioned and then of course the anode
that we removed and the screen itself
now if we look at the screen I thought
this is kind of interesting you can see
all these little defects those are on
the inside so what happened was when
this thing imploded a bunch of the
shards of glass scraped off some of the
phosphor the internal phosphor coating
on the screen and so even if the screen
was usable which obviously it never will
be again you'd have all these really
ugly defects that would make it
unappealing to to visualize so so that's
the CRT I'm gonna get this out of here
I'm going to clean up the rest of the
glass fragments that I have and I'm
gonna pull the new CRT clean it up a
little bit and get it installed
alright the CRT went in without an issue
and the traces show up as you can see
which is great but there's still a lot
of issues to deal with so let me show
you few of them so channel one when I
insert the probe into the calibration
port I get nothing
channel two I do get a signal which is
great you get the nice square wave there
so problem number one
is channel ones input is not working
problem number two is this channel
selection switch here is really light
what I mean by that is there's no
resistance when I use these other two
switches of the same type there's a lot
of resistance and there's really
discrete and definite steps when you
change positions and this one just kind
of almost free-floating
so there's some sort of mechanical issue
going on there that I'll have to address
and I looked around the board a little
bit more closely and I've discovered
that somebody has definitely done some
work on the power supply section so
we'll go to the power supply section
I'll show you what I found there and the
work that will need to be done okay
obviously we're looking at the back of
the board but this section here outlined
by this square is the power supply
section and you can see there's a lot of
redone so during saan here so you can
see there's this two rows of four up
here these are four resistors that were
for some reason replace they're
definitely not original
all of these solder joints need to be
redone we have a bank of four here that
were redone we have this one here the
one that's really concerning is this
jumper added right here no idea what
that's for that's gonna have to be
looked at we have a whole bunch of other
joints right here that have been touched
up we have some over here that have been
done and it looks like even that main
filter capacitor the seventy-five micro
farad at 450 volts was redone a little
bit added some solder to it or whatever
it's definitely still the original but
they added some solder to that for some
reason somebody touched up a whole bunch
of joints down here as well so a lot of
area for concern in just the power
supply section alone so all of these
joints are gonna have to be probably
completely redone let me show you the
top of the board with all of the
components
not as many problems to see from the top
if we look at the top so right here see
these four bigger resistors these are
the ones that were replaced those are
the ones that are not original the
there's three electrolytic that were
also replaced there's one right here one
right here and one over here not sure
why they only replaced three and why
those three otherwise the rest look to
be original there's some additional
solder on this jumper right here this is
the 100 volt test point and it's just a
jumper but for some reason there's a
whole bunch of extra solder on there and
you're not gonna be able to see it but
so there's a something mounted right
here so you can see this package right
here there's a transistor in there sorry
for the focusing the transistor that
drives the transformer here is mounted
here and then there's this block down
here and under there I believe there's
another transistor or two and I can see
a lifted trace sorry about the focus
again you're not gonna be able to see if
there's a lifted trace under here and
that is likely what that jumper was for
there's a whole bunch of issues to
address just in the power supply section
that's gonna fix some issues it's gonna
make this thing more reliable but it's
not gonna fix
it's definitely not gonna fix the two
issues and I spoke about earlier channel
ones input signal failure is gonna
probably come from this board here it's
got the attenuation for channel 1 and
channel 2 it's got the sweep selection
switch right here and all of the
necessary circuitry to deal with those
signals and then that switch that I
mentioned to with the likely mechanical
failure is also it's on this front board
here I've dealt with this type of board
once or twice in the past it's a pain
it's gonna be a pain to get this off to
get this front board off but it's gonna
need to be done so my first step is
going to be to remove and replace all of
these electrolytic sivan though three
that have been replaced I should be able
to find a fair replacement for this one
it's not going to be an exact match but
it'll be something very very close it'll
be a lot smaller than this older version
shouldn't mount to the board okay and
then I'm going to
as I said redo all of those weak solder
joints on the back and I'll keep updated
in front of the scope I have all of the
replacement capacitors lined up for the
a1 board the a1 board is the main board
on the bottom of the unit on the right
you can see I already removed the bulk
smoothing capacitor and the reason I did
this was to take a look at the lead
layout as I'm unfamiliar with this type
of capacitor so this is just an old
style and this bottom pin here which is
the smaller pin and a smaller lead is
the positive and the other three on the
outside are ground and so you can find
replacements for this online say on eBay
but I didn't want to buy used one so I
bought this new one it's a standard two
lead which would work just fine I will
fill in the unpopulated two unpopulated
ground through holes with solder should
work this one is slightly higher in
capacitance at 82 micro farad's versus
75 it's got the same voltage rating at
450 and obviously with a much smaller
size that's a benefit as well so I have
a lot of work to do I'm gonna get
started on this capacitor replacement
all the electrolytic have been replaced
on the main board the scope turns on
just fine and we still get nice clean
sweeps across the screen for both
channel 1 and channel 2 however we have
a new problem before the work at least
channel twos vertical deflection was
present I could see the calibration
signal with channel 2 but now I can't
so both channel 1 and channel 2 no
longer have the vertical deflection
their signals are not getting through to
the CRT so before I start looking at the
schematics to try and identify where
this problem might be I thought I'd look
at the power supply voltages first as I
replace all the electrolytes in the
power supply section so we'll start with
that
I don't expect there to be any problems
since the rest of the scope is
functioning but it's worth a check the
nice thing about this unit is that the
power supply voltages are all printed
nicely on the board and their test
points are very easily accessible so
let's start at the top I'll read them
out there should be positive 30 positive
20 9.8 that's okay
blow that 100 to positive 100 good way
down here positive 8.6 that's good
below that positive 5 good and below
that negative 8.6 and that's good so
write all of the power supply voltages
are present they're all very very close
to what they should be so I don't think
the problem is they're likely their
problem is going to be something to do
with this attenuator board here but I'll
take a look at the schematics to try and
identify where we want to start probing
all right finally troubleshooting time
so the first thing I want to look at is
channel 1 I want to look at this first
since this was a problem before I even
started working on this unit and so I've
done quite a bit of work I ready to
identify what the problem is you can see
that setup I have here so I have my B
and K 1477 scope hooked up that's what
I'm gonna probe with channel 1 or
channel a I have the Tektronix channel 1
hooked up to my B and K calibration
signal so when I hit the signal with the
probe I should get a nice square wave I
was able to identify that the attenuator
board is the problem with channel a I
checked the output of that board and I
got no signal so I know the problem is
somewhere on the board if we look at the
schematic the signal enters on the left
of the B and C it travels through a
whole bunch of different components
before exiting the right through that
connector that I mentioned and so
there's a few ways to do this you can
probe from left to right until you stop
seeing the signal you can probe from
right to left until you start seeing the
signal or you can start in the middle
and then if you see the signal you move
to the right if you don't see the signal
you can move to the left
I'm just gonna probe from left to right
so I want to look for some identifiable
and easily accessible components r2 is
pretty much where the signal enters the
board so let's look at our 2 here and we
have the signal there let's look at our
3 next we have the signal there as well
let's check the base of Q 13 and it's
there let's check the base of Q 18
it's there as well now we have this
resistor pack here it's gonna enter at
pin six here it's gonna exit pin five
here all right so a signal is making it
almost to the end of the line we're at
this custom Tektronix I see here it's
the u 30 and it's gonna enter pin four
so let's measure pin 4 and we have it
there and it should exit through a
differential signal out of pins 10 and
13 so let's look at pin 10 and you can
see we have nothing there and 13 we have
nothing there as well so it's likely
that that custom Tektronix IC is a
problem which is unfortunate because
it's of course decades old they haven't
made any new ones for decades you can
find a few here and there online but
they're fairly expensive for what they
are
that might be what we have to do the
only other thing you want to do to rule
out anything else other than the IC is
to check the incoming voltage to that so
if we look at the schematic again pin 6
has minus 8 point 6 volts coming across
our 34 and I measure there and that's
accurate and pin 14 should have positive
8 point 6 volts coming across the three
resistors there two resistors excuse me
and you should have something like two
and a half volts at that pin and that
was accurate as well so it's very very
likely that I see you thirty is the
problem with channel 1 now I'm not going
to order and replace that yet I want to
identify what the problem is with
channel 2 if the problem is also on the
attenuator board and the part is either
expensive or difficult to obtain it may
be easier and cheaper just to replace an
entire attenuator board so let's get
started with channel twos
troubleshooting I was unable to
replicate the problem I was seeing
earlier with channel 2 so I'm going to
ignore that for the moment I was able to
find a replacement for you 30 this is
the old one I was able to remove it
after quite a bit of effort and the new
one came in this genuine Tektronix box
and it was advertised as brand new so
we'll see if that works
before I test it I want to clean this
front faceplate here you can see all of
the exposed areas are much dirtier than
the unexposed where it was covered by
the knob so I'm going to clean all this
up really well I'm also gonna clean up
all of the previously reworked solder on
the back of the board mostly in that
power supply area and then we'll be
ready to test all right I clean the
front faceplate up pretty well it's
about as close to original as I'm gonna
get it I also cleaned up the back of the
board with all of the rework solder
joints and I put the cover back on and
clean that up as well so we're ready to
test it out and so I have a 1 megahertz
4 volt peak-to-peak sine wave at both
inputs and as you can see they both look
great
so I already checked most of the
frequency range on this scope at least
as high as my signal generator will go
that's 15 megahertz and everything looks
good I also went through a good section
of the voltage range as well my signal
generator goes up to I believe 20 volts
peak-to-peak and it looks good there as
well the last thing that I mentioned
previously of potentially wanted to
repair or replace was this channel
selection switch I mentioned that it was
loose there was no tension or no
resistance but I've tested it and as you
can see if I switch from channel 1 to
both the channel 2 back to both there's
absolutely no issue electrically with
this switch so I've already investigated
it a little bit this is attached to this
front PCB here and it's very very
difficult to access that really you have
to remove the front PCB it's mounted to
this bottom PCB and there's about I
don't know 30 or 40 solder joints to
remove and those are very difficult
joints to remove so I don't think it's
worth trying to remove that board to
investigate it any further since we've
already checked the voltages and
everything is looking really really
great I'm gonna call this one repaired
and I will see you in the next video
thanks for watching