DR #8 - Tektronix 2213A Oscilloscope Troubleshooting and Repair Part 1

Duality Repair
132 followers
632 Video Views·Mar 12, 2025
In part 1, I document the troubleshooting of a Tektronix 2213A oscilloscope with no channel 1 signal at the 100, 200, and 500 mV/division set-points.
hey everyone its Nick welcome back to
another duality repair video I have
another scope to look at tonight it's a
Tektronix 22:13 a 60 megahertz analog
scope and I got this off eBay again for
parts so there's something wrong with it
see it's got some pretty good condition
physically all the knobs and buttons are
there none of them are cracked or
missing or show serious signs of wear at
all well there's a few stickers on here
and some residue general scratches and
wear and tear on the case but nothing
too major the screen looks to be in good
shape
see there's a sticker on there that says
Pegasus lab that caught my eye it's kind
of curious so I looked them up it looks
like there's a company called Pegasus
laboratories in Florida does some sort
of chemical testing and R&D work so I
thought that was kind of interesting but
anyway there is something wrong with
this I'm not quite sure what it is so
let's turn it on and have some fun
alright the scopes plugged in but before
we power it up I noticed a few things I
guess just one thing it's trace rotation
here just an empty hole and I looked
further in there with a flashlight I
don't see any adjustment pod or anything
in there so I don't know if that's
missing or for some reason that's just
an option that was not included on this
eye that seems like it should be a
standard function so I'm gonna have to
open it up and see if that's missing at
some point I'd be kind of a bummer
especially if the trace needs the
rotation needs adjustment so one thing
to note before we power it up
so let's plugged in what's powered on
and see what we have so I'm on the both
channel 1 and channel 2 right now and
well the intensity sure works and
certainly only see one channel right now
so let's go to channel one that looks
like I'm seeing channel 1
huh so I'm toggling between channel 1
both and channel 2 and that beam isn't
moving at all so let's change positions
that's not right either
let's change oh man there you go okay
the time scale knob was really stiff I
thought it was just completely it seized
but probably just hasn't been cycled in
a while it seems to be working okay
density works really well so we do have
a beam but the vertical position doesn't
work
oops a vertical position for channel 2
works when I'm using the vertical
position for that channel so let's check
channel 1 alright it was working it's
needed to do the beam fine let's do both
yeah it doesn't look very good let's see
a little better
hmm on a tough time triggering once got
both beams up see they're both in AC
let's try the Chop there we go
okay now we're looking pretty good
channel one is adjustable vertically
channel two is adjustable vertically so
everything right now looking okay
question is are we gonna get a signal so
let's start with channel 1 and maybe I
mean I thought it was broken at first
but maybe it's just I've never seen a
probe adjust I guess it's just a pin
hole like that instead of the prong that
sticks out certainly see a change there
let's change the volts per division
thanks
a little ugly
that does not look like a square wave
should be 500 millivolts peak to peak at
100 or 1k Hertz 1000 Hertz
500 millivolts so we should easily see
something right here at 0.5 or smaller
interesting what happened they're not
looking too good channel 2 out see if we
get anything different on there again
I'll go to 0.5 since it's half a volt
peak-to-peak I'll go even smaller 0.2
channel 2 here
that's what I want to see yep channel 2
looks good
so looks like our problem is with
channel 1 not reading the signal
correctly or really at all so I point to
I should get the same signal as they do
here
I got nothing nothing at all well of
course not one channel 2 yeah it looks
bad that looks bad so we have a single
channel failure and other than that you
know everything looks to be good again
I'll have to figure out that trace
rotation but as far as major issues yeah
channel 1 channel one's no good so this
does have the service manual available
so I'll I'll pull that up and take a
look through it look through the
schematics and then see what kind of the
relevant schematics I might want to look
at and print those out and we'll take a
look and areas we want to focus on
alright so I'm glad I played around a
little bit more with the channel 1 volts
per division dial there before I opened
anything because take a look at this so
at 1 volt per division we have a nice
square wave and you know you get a
little bit higher and it gets a little
dicey but very nice at 1 volt per
division you go to point 5 and it looks
pretty rough right and later let's see
if this makes sense so 1 volt per
division it should be half a volt
peak-to-peak so if one division is 1
volt then we should have about half an
half a division so we're yeah that's
pretty close so what happens when we go
change the scale to half a volt it gets
pretty ugly its name with 0.2 but I
noticed as well its 0.1 and you go right
there so you can still see it you can
see the bottom of the square right here
you just can't see the top because the
the scale is so big and I wonder if you
can see it well we can do this too we
can move it up and down yeah so you can
see that there and if we go down to 20
20 millivolts per division you can see
it there it's getting a little rough and
10 I'm not seeing much there yeah you
can see it there too so it kind of just
looks like we might have a dirty dial to
the rotary encoder there so go back to
one and of course it's no it's not
reading nicely at one oh that's go say
it's messing with the fine adjustment
see if that'll help here really so I'm
not gonna worry about the schematics at
this point I'm gonna open this up and
see how easy it is to clean this uh just
knob here and the rotary encoder that
might be our only problem while I'm in
there I'll take a look at this trace
rotation and see if it's broken off or
it's not supposed to be in there or what
the heck is going on there so
yeah we'll figure it out here before I
move on to opening this I think it's
worth noting that I figured out there is
a pot back here for the trace rotation
and it's obviously working they just for
whatever reason decided to make it
recessed by oh that much
well three quarters of an inch to about
an inch so it is back there it is
working so I'm not gonna worry about
that just gonna worry about these
cleaning these dials off so let's get it
open
the unit's out of the case let's take a
quick look around so we have our power
supply section hidden in here it's
definitely better protected than the
22:05 I just worked on they have this
nice shielding case I'm sure both for
protection as well as for noise
mitigation then of course you have the
seer CRT up here and the high voltage
cable leading from the power supply
section over here we have the front
board so your front panel board that
will distribute your signals like
position vertical position horizontal
position triggering things like that and
then down below here we have our
attenuation and suite board so this is
going to relay our signals for our
channel 1 channel 2 vertical deflection
as well as our our time scale through
these so these are actually for the fine
tune adjustment so you can see I can
rotate this one that's the fine tune for
channel 1 channel 2 vertical deflection
and then here's the fine tune for the
time scale and then the main vertical
deflection adjustment is going to be
under here
and you can see the main vertical time
scale is right here but I know that
these are slightly different so we'll
have to pull this cover off to take a
look we also have the main board on the
bottom it's one big giant board and of
course Tektronix makes it nice and
serviceable pull it out and it's it's
all it's all exposed right there very
nice very nice so I don't think I'll
have to be doing any work on the main
board here at least not yet
get started on cleaning cleaning the
channel one vertical deflection so I'll
take this cover off and we'll see what
it looks like underneath alright so in
order to expedite the troubleshooting
process let's go through this channel 1
volts per division knob here let's go
across the entire dial and identify and
characterize exactly what the problem is
so that we don't waste time looking at
parts of the circuit that aren't
relevant to the problem so I am fully
counterclockwise and let's hook it up to
the signal so with you know it's such a
small signal it's it's not gonna look
very well with the trigger but it is
showing just fine at the 5 volts per
division here at the x 1a probe so let's
turn the dial clockwise and and see what
we get each time so at 5 we're good at 2
we're also good again the triggering
isn't great there we go it's still not
triggering wonderfully but you can you
can definitely see the square wave so I
5 were good at 2 we're good at 1 volt
we're still good ok when we get to the
point 5 volts this should be a really
nice looking square wave okay we don't
have that there point 2 volts
same thing we should be able to see it
it would probably go off the screen in
one or both directions but we would be
able to see it easily and obviously it's
not there so 0.5 is bad
my two is bad it's checkpoint 10.1 also
bad right so all three of these 0.5 0.2
0.1 are bad let's go to 50 mil 50 mil is
good so it's huge but it's they're
seeing it very nicely let's go to 20 20
mil also there and then 10 also there I
can't even scroll to it but you can see
the vertical lines there 5 same ten
cities dropping off there there we go 5
mil same 2 mil you can also still see it
low intensity but they're all the
present their turn intensity back down
our issue is going to be with point one
point two and point 5 volts per division
only and that will come in handy and
it's extremely relevant for the
troubleshooting process so keep that in
mind
Oh point one point two point five all
right I have the cover removed from the
attenuator sweep board and so you can
see these two rotors here are the
controls for the channel 1 channel 2
volts per division adjustment so let's
zoom in here on channel ones rotor so
you can see what's going on so what's
happening here is that you can see along
this rotor there are these little nodes
jutting out and what these do is as you
rotate they either press down or they do
not press down on these little Forks and
so when the node is not pressing down on
that fork there's tension on the fork
that allows it to be raised and to be
elevated off of the PCB as you rotate
and the node pushes it down it makes
contact with the PCB so this is what's
allowing for different circuits to be
activated as you rotate it now looking
at this two things a I don't see any
sort of corrosion or carbonization any
build-up on any of these contacts I
don't think this is a contact related
issue and then the other thing I notice
is I was worried about maybe we love
lost some of the spring tension in some
of these little Forks here but I checked
continuity and all 16 as they were down
and they all made they were all
continuous so I don't think that is the
issue I think what the issue is is part
of the circuit itself that's involved
with the three volts per division set
points I think something in those parts
of the circuit is as faulty so one thing
I had to do was the reason I went over
earlier which set points where faulty is
so that I can help categorize which of
these forks may be bad
and then which part of the circuits they
correspond to so so as you'll see this
is fully clockwise this is 5 C here
fully counterclockwise this is 5 volts
per division as I rotate through to 2
volts per division and then 1 volt per
division only one of these Forks changes
in between each all the rest are the
same so within each area of adjustment
only one fork changes between the two so
now when I change from one volt per
division to half a volt which is our
first problem area you get several of
these other Forks change
ok and so now when I change within this
range from 0.5 to 0.2 again we only get
one fork change and from point 2 to
point 1 again only one change so then
when I go back down to another scale
another range which is 50 millivolts
I get all these Forks changing again so
these are grouped in ranges and there
are three ranges of three set points so
I did a bit of work there are 16 Forks
along this control here and the only two
that are common between our faulty range
of 0.5 0.2 0.1 volts per division and
uncommon - all of the rest of them is
fork 6 and 49 ok so I look really
closely at these I didn't see any
buildup on
and I verified they are making contact
when they're supposed to with the PCB
below it so what I think we need to do
is pull the schematic for this sweep
attenuator board and look at the
components that are in the circuit
making contact with fork six and fork
nine and then we have something to go
after you know maybe there's a resistor
in line or capacitor in line something
that's not behaving correctly which is
why we're getting basically no signal in
this range so I'll pull up the schematic
we'll take a look I have the attenuator
board schematic here and specifically I
have it zoomed in on the channel one
portion of that schematic and you'll see
I have some things highlighted here I'll
explain that in a second but what's
important to identify is that this whole
section right here is the rotary switch
and so if we look on the board here
that's this okay and yes I did remove
the board I'll talk about that in a
second so let's zoom in on that a little
bit so we started all the way on the
left we can see that we have our
millivolts per division labeling so two
five ten all the way up to five volts
and then I have 100 200 500 millivolts
per division highlighted because those
are the three set points that we were
having issues with it's in all the way
this works is you can see each of these
columns is labeled 1 2 3 etc all the way
up to 15 at the end and so in the
previous segment I described I said
there were 16 Forks or contacts but I
either miss counted or their labeling or
their counting two contacts together as
one and so it's one of those two things
and so you can see that these dots here
in each column the dot represents
contact and no dots represent no contact
so when the when the nubs on the rotary
dial there are pushing down on the
contacts that's represented by these
three dots or these dots I should say
three dots in this case
all right and so I spoke last time about
commonality between these three set
points and which contacts have
commonality and so you zoom out a little
bit we can verify that the only two
contacts that are common to these three
set points are these two right here the
ones I have highlighted you can see this
one there's they're not even making
contact this one they're making contact
but so are the rest of these set points
this one the same thing they're making
contact with the rest of these set
points and etc the only ones that are
common are these two little two men back
on these so you can see contact five and
contact eight are the only two contacts
that utilize the 100 200 500 millivolts
per division and the previous segment I
think I called them contact or fork six
and nine and so since I miscounted or
whatever they're labeling them
differently I was just off by one so
where do these lead these lead up to
this attenuator this 10k attenuator
attenuator to here and so basically
attenuator is just two capacitors you
can see there's two knobs with c1 and c2
c11 excuse me in c12 and so these are
just two capacitors inside this
attenuator here and this is a custom
package manufactured by Tektronix and so
first thing I did before I remove the
board was I adjusted c11 basically all
the way in one range and all the way to
the other didn't change anything I
adjust the c12 one range all the way to
one range of all the way to the other
and it didn't make any difference I just
said they're both at the same time
didn't show any difference to the signal
so I don't believe I don't believe that
adjusting this is going to fix the issue
and so unfortunately I think this
attenuator is what's causing the issue
basically these two capacitors in line
in this one package the other thing I
did was I tested I tested for shorts or
opens between the capacitors and to
ground I didn't see anything different
between this attenuator here and the one
down on channel 2
so this is channel twos here that's the
way they have it spread out so
attenuator fifty-two is for Channel two
I compared continuity between all the
pins on this attenuator on channel 2 and
channel 1 there was no difference so
what I'm thinking is one or both of
these capacitors either have there's
some sort of issue with this attenuator
it has to be because there's nothing in
line but the between contact 5 and
contact 8 and the attenuator
there's no resistors capacitors any
other passives in between here and this
is the only difference between our 100
200 500 millivolt
and the rest this is the only difference
there's no difference at all the rest
utilize you know similar parts of the
circuit there's all there's commonality
between other parts of the circuit
between these 3 set points and the rest
so it has to be this attenuator that's
causing our issue so let's take a look
at the board I removed it and so we have
channel 1 here
channel 2 you can see these resistors
off here on the right those were they're
soldered on the back and then they were
soldered to the input connections well I
just D soldered I'm gonna remove the
board so if we're looking at the board
here we have the 2 attenuators right
here and so look a little closer you can
see 10 X on the top one and 100 X on the
bottom one so this is our attenuator
that we're gonna have a problem with and
they have the Tektronix part number
right down there 307 - 10 13 - 0 0 and I
looked it up yeah you know they're they
have references to them all over the
internet but none of them not nobody
shows them in stock except for one
person in greece on ebay and it was
$7.00 so i bought it that's the only one
i could find
$7.00 plus 8 dollar shipping from greece
yes greece I think the 15 dollars is
well worth putting into the scope since
this seems to be one or solely the only
issue on this scope I think it's well
worth it with the price I paid for it
putting
fifteen dollars into what I'll be able
to get the scope up and running and so
since its from Greece it's going to be
two to three weeks before I get this
thing in which is why I'm making this
part one I'll make part two when that
comes in when it comes in I'll remove
this I'll install that will verify that
our issue is gone on the 100 200 500
millivolt per division set points and
then we'll also verify the functionality
the rest of the scope if there's
anything else wrong with the scope will
attempt to address it at that point so I
will be back with part two when I get
this attenuator in and we'll start by
installing it in part 2 thanks for
watching we'll see you next time