hey everyone its Nick welcome back to
another duality repair video today I
have a Tektronix 2205 20 megahertz dual
channel oscilloscope just got this in
from ebay I bought it for the very
reasonable price of $30 I believe the
issue was it's got some issues with the
vertical deflection so let's power it up
I have not powered it up this will be
the first time it's plugged in let's
power it on and see what it's what's
going on with this thing
can we have a beam it's obviously good
and yeah
sure you can see that yeah so with no
input we have some noise in the vertical
deflection so that's what they're
talking about
I've never worked with one of these
Tektronix 2205 so bear with me as they
figure out where all the buttons and
knobs are
I'm not crazy about the focus so this is
channel 1 let's see what channel 2
produces let's go to the same volts per
division right now I'm at one it's got
it too so if we go to both yeah they
have the same thing mmm so obviously
there's not a problem with a single
channel but something that's common to
both channels so that's what we're gonna
have to figure out let me change the
horizontal here keeps in line pretty
well
and see it all the way up
curious what it will look like if I am
if I have something hooked up so I hook
it up to the Cal signal maybe when I go
back to channel 1 let's go to channel 1
and where is it
let's just put a probe in there and hook
it up to the Cal signal and see if see
what we get curious
you can see so that was the ground DC
and AC both look like garbage I doubt
I'm gonna see anything on this
yeah it doesn't look good DC not some it
just looks bad doesn't look good my
channel too quick how much going on
there
see that change a little bit well it's
just adding it to that noise great ugly
though obviously some noise going on
we'll have to pull up the service manual
take a look at what could possibly be
causing this first guess would be
capacitor a capacitor related issue on
the vertical deflection circuit or maybe
the power supply can be power supply
itself with some noise not sure yet so
let's pull up the schematics and take a
look before we start digging into the
schematics and looking at the vertical
amplification circuit I pulled this up
from the service manual it's got the
power supply test measurements here so
you can see this little table here on
the right and it's got six voltages to
measure negative eight point six
positive five positive eight point six
positive 102 positive 2 O 5 and positive
22 unregulated five of these six are on
the main board and I've pulled their
locations up here service manual and of
course this is a scanned copy that's
what was available online so it's very
difficult to read the read everything
they can legible at all so I I wrote
down each position you can see the
positions here they pretty much line up
w-9 $89.99 91979 92007 the middle here
so I should be able to find those pretty
easily this sixth one 22 ma related we
find on a time base 10 you meter board
and that's right at the top so that
should be fairly easy to find
depending on where these boards are
located so we'll go after these first we
will make sure that they are within
specifications and then we'll see if
there's any ripple on any of these
specifically I believe the vertical
board uses negative and positive eight
point six and the positive five so these
three we're on
pay particular attention to might as
well check the other three as well
because they're they're all critical so
I will get this thing open and we'll
take a look at these voltages all right
I was able to remove the unit from the
case and I haven't flipped over on its
side exposing the main board and so
let's just go through these test points
one by one starting with 989 this should
be negative eight point six volts and
we'll check DC first so 989 is up here
looks pretty good
negative eight point six it's 291 which
should be positive five nine ninety one
is here 5.06 good 987 should be positive
eight point six nine eighty sevens here
it's pretty close eight point seven
expect for that is two eight point eight
six so that's good okay let's go to nine
eighty four which is 102 volts so I need
to change my scale here 984 102 point
nine hundred and three volts the spec is
plus or minus 3 volts so that one's good
nine eighty five should be plus two oh
five volts yeah change skills again nine
eighty five to ten to eleven that's the
upper end of its spec upper end is to 11
point one so we may want to adjust that
one it's probably for a CRT voltage
there and that final voltage is on the
attenuator board so we'll have to look
that one later
that one's unregulated anyway and I
don't believe this has any effect on the
vertical so I may skip that for now but
let's go back and check AC on all these
test points that's really what we're
curious about right so let's check AC
starting with w 989 the negative eight
point six points 989 checking for AC
not a good 991 positive 5 volts very
critical nothing 987 should be + 8.6
chekc on that one
nothing and then we'll check 102.9 84
which is 102 might be a little bit on
that nope nothing
and 9 85 here
practically nothing okay mm-hmm I think
it's safe to say that we don't have a
power supply problem as in a filtering
problem causing Ripple we traverse its
outputs specifically the vertical boards
so now I'll have to dig into the
vertical amplification circuit and take
a look see where we want to poke around
there so I'll be back when I have that
circuit print out here is the schematic
for the vertical amplification circuit
I understand it's pretty difficult to
read not only it's very small but it's
as I said it was scanned it's a scan
version so it's it's almost eligible but
we can get the general gist of what
we're looking at here so you look at the
top here this is the channel 1 input and
the preamplifier the bottom is channel 2
input and preamplifier and then in the
middle you have your channel selector
switch I'm not really concerned with
these areas since we have noise on both
channels there's hardly any commonality
between them here at all when I'm going
to start where I'm going to focus is in
this middle section here which is the
output amplifier driver circuit and
really there's not much common between
them there either but there's a little
bit so I have a zoomed in picture of
this so we have our signals from channel
1 coming in here on the left and the top
and channel 2 on the bottom each
obviously is fed into a transistor q2 l
3q tool 2 and then they go their
separate ways
channel 1 up through this transistor and
out channel 2 down through this
transistor and out well where do they
have commonality right here in the
middle mm-hmm so they have this resistor
R 2 O 4 this resistor R 2 15 and this
capacitor C 2 15 I believe may be 16 so
these are items of concerned
specifically that capacitor all right
anything that is common between the two
channels is of concern where I'm going
to test
first is on these test points so there's
a test point here and they're actually
shared W 224 and W 225
I'll find I was able to find W 224 so
I'll check there first so let's take a
look at at that point and that's gonna
be right there on the board you can see
W 224 right here which is gonna be right
there so I'm gonna power the unit up
it's not plugged in yet I'm gonna power
it up
don't power it up now and we'll see if
we get any noise on that mm-hmm test
point if so well we can rule out
portions of the circuit based on what we
find we go it's powered on I'm gonna
give it a second make sure we have a
beam yep I have noise on that beam so
let me probe the test point like that
noise looks identical to what we have
your not identical but that's nasty
looking
No
okay that means we don't really have to
go any further into the amplification
circuit we have to go backwards and I
said specifically that capacitor was of
concern because that's common between
the two I might check there first
so let me show you that again so again
this is the output amplifier driver
circuit right so here's the main circuit
for the entire circuit right look in
here with the output amplifier driver
and I'm testing here and I'm getting
noise so that means I don't have to
worry about looking post this test point
it's not happening we're not getting
noise introduced in the vertical an
output amplification circuit we're
getting noise introduced probably here
or maybe somewhere back but I'm gonna
focus still on the output amplification
driver circuit which is again here again
I'm testing here I have noise on these
points so we have to work backwards so
right now I'm gonna try and find this
capacitor it's 215 or soot 216 difficult
to tell and I'm gonna test on either
point on either end of that capacitor
and see if this capacitor is what's
causing our issue so let me try and find
that alright I was able to locate C 215
so I'm gonna start probing see if I get
a signal on either side here
nothing on that side some strange noise
on that side
so I'm back on the test point just to
get it
whoops locked in here there we go now go
back to see 2:15
okay
so this is my input on one side
and
that's what I get from the other side of
that it's nasty you know the resistor
that's in line with that I'm not good
either so no it doesn't look identical
to this when you have an input like this
to the capacitor and an output like that
I'm gonna say that capacitor is faulty
so I flip the unit over we'll take a
look at it see if see if it physically
looks in bad condition or not regardless
I'll likely want to replace that and see
if that makes a difference
pull the capacitor out and I'm
definitely second-guessing myself at
this point the capacitor is ceramic it's
not a electrolytic I don't think that
thing failed so I went back to the
schematic and in here here's the
capacitor in question again no noise on
this side and then the noise is
introduced on this side of it so I
assume this capacitor was bad but I
noticed that our negative 8.6 voltage
input is coming in right here so now I'm
thinking about the power supply again so
what I'm gonna do and I know I did it
earlier is I'm gonna check those test
points again he was using the multimeter
on the AC mode that's not necessarily
gonna pick up the noise so I'm gonna try
it with the scope I have the scope
hooked up there I'm gonna power the unit
on so let's take a look
so let me run through the test points
again we're gonna find the Sheep test
points on here I'm gonna start at the
negative eight point six right here
negative eight point six is test point
nine eighty nine so let me check that
huh so look at that good deal annoys
there obviously it's not regulated very
well let me test check the other one
plus five volts 991 let's check that
well plus five has very similar thing so
this should be a straight line obviously
if it's filtered properly coming into
the unit let me trace back a bit I can
see the traces on the board so I'm gonna
look at the negative eight point six
rail and go back to
[Music]
see so this is negative 8.6 this is the
inductor that's in line with it we still
have noise here a lot of noise on that
ending inductor in dr. actually pulls
out a lot of the noise
[Music]
and then here's the capacitor in line
okay so this is again the test point for
negative eight point six let me go to
the capacitor that's in line with them
so that's one leg of the capacitor and
that's the other way so we're just not
filtering and and it makes more sense
because these are electrolytic s-- so
most of the board also this main board
contains ceramic capacitors power supply
however power supply section just on the
main board contains several
[Music]
so here's the mainboard
you can see there's tons of capacitors
but a majority of them are ceramic which
aren't gonna feel very often come over
the power supply section you see the
main transformer there the transformer
and then you have your capacitors
these are electrolytic s-- these are
subject to fail especially if they're
original then but in this unit for
30-plus years so I think what's gonna
get rid of those noise is to replace all
these electrolytic mmm thankfully
there's not a ton there's probably 10 in
there and I think a lot of them are the
same size and rating so that shouldn't
be too bad so I know I don't have all of
these spares so I'm gonna have to order
some of these I'll get some new ones in
we'll replace them and we'll see if
we'll see if that makes a difference I
think it will so I'm gonna order these
up and I'll be back when I put them in
old caps are out no caps arrived What's
in stone
it's good
there's one mine where to go
all right they're all in I'm gonna take
a few minutes to make sure that I have
them all in the correct orientation
because none of them are actually marked
on the silkscreen only one of them where
is it up here it's the only one I can
find with the marking on it I'm pure
positive none of the rest have them for
some reason
so the designation is the through-hole
shape is either circular or square and
as you can hopefully maybe tell up here
the square is positive in the circle is
negative so I'll have to go around make
sure it'll be easiest just to test from
ground to each negative and make sure
that they're all in the correct position
and then verify have the correct
polarity on the capacitors
if those all look good I will clean the
bottom of the board up and we'll power
it on see what we get I verified the
correct polarity the caps I clean them
up a little bit clean the board up a
little bit I should say and I have the
scope plugged in and ready to turn it on
let's see if we have any magic smoke
certainly hope not
all right it's on so let's take a look
at that test point with my scope
it was W 989 up in your upper right
corner there the negative eight point
six volts this one that's directly tied
into the vertical circuit so let's probe
that it's here we go hopefully we don't
get that noise that we got before now
obviously I could just look at the
screen on the scope and see if the noise
is there I want to check this out first
hey that looks pretty good so I'm on
test point W 989 and rewind just a
little bit to get a refresher on what
that noise used to look like and I'm
checking both sides of that cap cap 989
which is filtering the negative eight
point six volts and it looks good let me
check some of the other test points they
also had some noise on them this is the
plus five point two volts let me get it
on there they're a little bit oh that
didn't sound good that didn't sound good
something just popped there I'm gonna
power that down yep something just
popped oh boy
well let's I would unplug it and turn it
around take a look
Baumer so the two caps in the middle
there you can see the one on the right
pretty clearly right here it's bulge
that one blue and then the one just to
the left of that blue as well the other
eight looked to be in good shape I so
this is directly both of them are
directly attached to two out points I
was testing so I may have inadvertently
shorted something so I'm gonna replace
those again will power it up and we'll
see if we'll see if everything's okay
it's not I'll probably to dig a little
bit deeper I did verify the orientation
is
the polarity as well as the the capsizes
and ratings themselves are correct so
I'll replace those I bought some few
extra spares and we'll see if that works
any better all right I removed their two
caps and I blew these things to
smithereens you can see the one on the
left is a little bit worse but they're
both blown that's after being installed
and powered on for maybe 30 seconds so
you can see the one on the left started
leaking out of the bottom tore it maybe
it ran down the side but it this one
just blew really bad and this one as
well the other ones look fine and I'm
pretty confident that these are the only
two that are gonna be at fault and I'll
show you why they found out what went
wrong so this is a clip of the power
supply circuit and this is the and i-i-i
i understand that the the resolution on
this drawing is is terrible but again
this was a scanned copy so i have to
work with what i have so this is the
positive 8.6 and then negative 8.6 rails
and so mm I made a mistake
I don't percent my fault these two
capacitors are these two capacitors and
if you look at the orientation the
positive is to ground in both cases and
they make sense because this is the
negative rail so that was a complete
oversight on my on my part 100% my fault
I I wasn't thinking about that negative
rail when I was installing the caps I
just wanted to get them installed and I
tested tested each pin of the of the
through holes for the negative and I put
the negative leads of the capacitors in
each but these two I obviously needed to
switch around so I've removed these I
installed two new ones with the correct
orientation with positive to ground
and then I'd have to be negative 8.6
rail I just hope I didn't damage
anything permanently on the board or any
other components it's possible that I
did
we'll find out so you know part of being
electronics technician is making
mistakes but to become a good successful
electronic repair technician you have to
acknowledge and learn from your mistakes
so I assure you I haven't made that
mistake before and I will not make it
again so let's move forward with the
testing do the exact same thing I did
before I'll power it up
we'll check for noise and I'll leave it
on for a few minutes to see if I hear
any fizzling I sure hope I don't again I
may have fried something else we'll see
I hope not
so here goes nothing I'm gonna power it
on right now no immediate smoke
so let's take a look at these two caps
let's start with 989 all right it's
there's very minimal noise on there
little bit that may be expected
especially compared to where we were at
before and then I'll check the other cap
it's a bit of noise on there a little
concern about that let's tech check the
actual test point w 989 which is the
negative 8.6 a little bit of noise
little bit we definitely had a lot worse
noise before change the ya know it
doesn't look like it did before so we'll
see check a few other test points this
is the positive 5.2 that's the same so
that may be a standard amount of noise
Emily just powered on for a few minutes
and make sure I don't hear anything
sizzle and if not we'll be back to look
at the look of the signals and see if
that noise has gone if we have more work
to do
here's a 10 kilohertz signal going into
both channels looking pretty good again
a little touchy on a triggering so that
I might need to look at that a little
bit but that could be a source issue I
don't have a generator at this moment
let's cut that so I have a 10 kilohertz
sine wave going into both channels and
it's looking pretty good
the triggering looks a little shaky but
I mean you can see the sine wave pretty
stable were there in both channels the
they overlap very well so I would say
there's no adjustment needed as far as
uh any amplitude adjustment the gain on
either channel looks pretty proportional
pretty equal to each other so that's
good
I don't think I can go any higher as far
as any higher of a frequency because I
don't have a signal generator at the
moment what I'm using right now is a 10
kilohertz tone file from YouTube via my
laptop I'm plugging in a simple audio
cable to the headphone jack and then
using Oh disconnecting them using the
probes to connect up to those and that's
what's giving me my signal if I go to
much higher
I think the capabilities limitations of
the computer itself show that the it's
just a lot of noise on here and this is
a 20 megahertz scope I highly doubt that
the scope can't handle 50 kilohertz I
think it's it's more of my source so I'm
gonna call this one looking pretty darn
good right now we have this to get rid
of the noise with the caps and each
signal is pulling up the each Channel
is pulling up a trace very well
everything seems to be working pretty
well
tensity gives out and maxes out focus is
adjustable no can see that clearly out
of focus there and there I'll have to
play around with a few others the volts
per division obviously working trigger
means tough to trigger there
and our time scale knob is working
horizontal position I guess I haven't
checked that yet that looks pretty good
I checked both vertical positions I am
an auto trigger you can put it in the
norm and it lets me manually trigger and
I should be able to you know feel the
trigger obviously I prefer auto can
change the slope so you'll see the
trigger changes based on the the rise
versus the fall of the slope so that's
working everything seems to be working
pretty well on this I'm gonna call this
sculpt this 22:05 20 megahertz Tektronix
scope fixed and I'm gonna get it cleaned
up I might clean up these dials verify
everything's working again and then
clean it all up externally get it back
in its case and then make room on the
bench for the next scope I have a 22 13
I just got in so I'm ready to work on
that as soon as I can so we'll see on
the next one thanks for watching